Original Article |
1 Department of Angiology & Blood Coagulation "Marino Golinelli", University Hospital S.Orsola-Malpighi, Bologna;
2 Department of Hematology, S. Bortolo Hospital, Vicenza and
3 "Angelo Bianchi Bonomi Hemophilia & Thrombosis Center, Department of Internal Medicine, University & IRCCS Maggiore Hospital, Milan, Italy
Correspondence: Cristina Legnani, Department of Angiology and Blood Coagulation "Marino Golinelli", University Hospital S. Orsola-Malpighi, via Albertoni, 15, 40138 Bologna, Italy. E-mail:cristina.legnani{at}aosp.bo.it
| ABSTRACT |
|---|
|
|
|---|
Design and Methods: VIDAS D-dimer Exclusion (bioMerieux), Innovance D-DIMER (Dade Behring), HemosIL D-dimer HS (Instrumentation Laboratory) and STA Liatest D-dimer (Diagnostica Stago) assays were performed in plasma aliquots sampled 30±10 days after cessation of vitamin K antagonism in 321 patients enrolled in the PROLONG study.
Results: During the follow-up without vitamin K antagonism, 25 patients had recurrent venous thromboembolism. The cut-off levels of the quantitative assays giving results most comparable with those of the qualitative test were: VIDAS = 800 ng/mL; Innovance = 800 ng/mL; HemosIL HS = 300 ng/mL; STA Liatest = 700 ng/mL. When the effect of the patients age (
70 vs. >70 years) was analyzed, it was found that only in younger patients was the rate of recurrence of venous thromboembolism significantly higher in patients with abnormal D-dimer levels. However, using the quantitative assays and age-specific cut-off levels it was possible to determine statistically significant hazard ratios also in elderly patients (VIDAS = 600 and 1200 ng/mL, Innovance = 500 and 900 ng/mL, HemosIL HS = 250 and 450 ng/mL, STA Liatest = 700 and 1000 ng/mL, in patients aged
70 and >70 years, respectively).
Conclusions: Quantitative D-dimer assays may provide information useful for evaluating the individual risk of recurrent venous thromboembolism. They seem particularly advantageous since they allow the selection of different cut-off levels according to the age or other characteristics of the patients.
Key words: anticoagulation, D-dimer, recurrence, risk factors, venous thromboembolism.
| Introduction |
|---|
|
|
|---|
Earlier prospective studies in patients with VTE9–12 demonstrated that D-dimer levels have a strong predictive value for the occurrence of subsequent episodes. These studies suggested that D-dimer measurement may have a role in gauging the duration of anticoagulation in patients with VTE. The recently published PRO-LONG study,13 a prospective, randomized, multicenter study investigating patients after a first episode of symptomatic unprovoked VTE, showed that patients with abnormal D-dimer levels after withdrawal of anticoagulation had a higher rate of VTE recurrence than those with normal D-dimer levels and benefited from resumption of anticoagulation.
Since PROLONG was a multicenter study and D-dimer measurements needed to be performed with the same test in all participant centers, a qualitative D-dimer assay was chosen (Clearview Simplify D-dimer by Inverness Medical Professional Diagnostics). This test can be performed on citrated whole blood and the result (negative or positive) evaluated visually. In principle, however, the accuracy of D-dimer assays may be enhanced by using quantitative tests with appropriate cut-off values.
The present study retrospectively analyzed a sub-group of patients enrolled in the PROLONG study. Its main aims were to assess the performance of four quantitative D-dimer methods to predict the risk of VTE recurrence and to verify whether the adoption of different cut-off values may help to distinguish better between subjects with higher or lower risk of recurrence.
| Design and Methods |
|---|
|
|
|---|
At the visit scheduled 30±10 days after interruption of vitamin K antagonist treatment, besides performing the Clearview Simplify D-dimer assay, the participating centers were also asked to collect plasma aliquots for further centralized analyses; a total of 386 plasma aliquots were collected and stored in 14 participating centers. The present analysis was centrally performed using plasma samples from 321 patients, 240 with normal Clearview Simplify D-dimer results and 81 with abnormal results; the 65 samples collected from subjects with abnormal Clearview Simplify D-dimer results who had been randomized to resume treatment with vitamin K antagonist were excluded.
This analysis considered the same outcome as in the PROLONG study: objectively documented recurrence of DVT and/or PE.13
Blood sampling and D-dimer tests
Blood was collected from the antecubital vein into 0.129 mmol/L trisodium citrate; plasma was prepared by centrifugation for 20 min at 2000 g at room temperature. Platelet-poor plasma was distributed into coded plastic tubes, snap-frozen and stored locally at –70°C. Frozen aliquots were then sent by courier to the coordinating center in dry ice and stored at –70°C. The Clearview Simplify D-dimer assay (Inverness Medical Professional Diagnostics, Bedford, UK and Luisville, Colorado, USA; kindly provided by Instrumentation Laboratory, Milan, Italy) was performed locally in each center on fresh citrated whole blood.
In the present study, we report the results obtained with four quantitative D-dimer assays which were performed centrally in the laboratory of the co-ordinating center at the end of the follow-up of the PROLONG study. These D-dimer tests were performed (after thawing the frozen aliquots for 5 minutes in a waterbath at 37°C) by technicians unaware of the clinical characteristics of the patients. The following assays were performed: (i) VIDAS D-dimer Exclusion (bioMerieux, Lyon, France; kindly provided by the manufacturer) on a VIDAS apparatus (bioMerieux); the calibration curve stored in the memory for each lot was reset every 14 days using a provided calibrator; (ii) Innovance D-DIMER (Dade Behring, Marburg, Germany; kindly provided by the manufacturer) on a BCS instrument (Dade Behring); calibration was performed using the calibrator included in the kit and was carried out monthly or when a new lot was used; (iii) HemosIL D-dimer HS (Instrumentation Laboratory; kindly provided by the manufacturer) on an ACL TOP instrument (Instrumentation Laboratory); calibration was performed using the D-dimer calibrator included in the kit and was carried out monthly or when a new lot was used; and (iv) STA Liatest D-dimer (Diagnostica Stago, Asnieressur-Seine, France; kindly provided by Roche Diagnostics, Monza, Italy) on a STA Compact instrument (Diagnostica Stago); a calibration curve stored in the memory for each lot was used.
The results of VIDAS D-dimer Exclusion, Innovance D-DIMER and STA Liatest D-dimer are expressed in ng/mL of fibrinogen equivalent units, since these assays use crude plasmin digested lysates of cross-linked fibrin clots as calibrators. The calibrator used in the HemosIL D-dimer HS assay, however, consists of purified D-dimer fragments obtained from a plasmin-digested clot and so the results of this assay are expressed in ng/mL of D-dimer units. Since one D-dimer unit corresponds roughly to two fibrinogen equivalent units, the values obtained with the HemosIL D-dimer HS are expected to be about half those of the other assays.
Statistical analysis
Continuous variables are presented as median (range). Hazard ratios (HR) and 95% confidence intervals (CI) were calculated using a Cox proportional-hazard model. The SPSS statistical software package (Version 11.0, Chicago, Ill, USA) was used for data processing.
| Results |
|---|
|
|
|---|
|
Table 2 shows the results calculated for each quantitative D-dimer assay using the cut-off value indicated by the manufacturers for diagnostic strategies regarding exclusion of VTE in symptomatic patients. The results, also calculated at different cut-off levels (see Figure 1), were similar for all the evaluated tests. By increasing the cut-off level fewer patients had abnormal D-dimer results and the percentage of VTE recurrences in these patients increased; however, the absolute number of VTE recurrences in this kind of patient decreased. In patients with normal D-dimer levels the absolute number of VTE recurrences raised in step with increasing cut-off values; however, the percentage of VTE recurrences did not change significantly. Indeed, the negative predictive values at the lowest and highest cut-off levels were, respectively: VIDAS D-dimer Exclusion = 97.4% and 95.1%; Innovance D-DIMER = 95.7% and 93.8%; HemosIL D-dimer HS = 95.8% and 94.4%; and STA Liatest D-dimer = 95.9% and 94.4%. Table 2 also reports the results obtained using as the cut-off level the one giving results most comparable to those obtained with the Clearview Simplify D-dimer assay: VIDAS D-dimer Exclusion = 800 ng/mL; Innovance D-DIMER = 800 ng/mL; HemosIL D-dimer HS = 300 ng/mL; and STA Liatest D-dimer = 700 ng/mL. We also calculated the incidence rate of VTE recurrence per 100 person-years in patients with abnormal and normal D-dimers results using different cut-off levels for each quantitative D-dimer assay (Figure 2). At the cut-off levels providing the results most comparable to those obtained with the Clearview Simplify D-dimer assay, the incidence rates in patients with abnormal/normal D-dimer levels were: VIDAS D-dimer Exclusion = 11.6/3.5 per 100 person-years; Innovance D-DIMER = 12.2/2.9 per 100 person-years; HemosIL D-dimer HS = 11.4/3.9 per 100 person-years; and STA Liatest D-dimer = 13.2/4.8 per 100 person-years.
|
|
|
70 vs. >70 years). For quantitative D-dimer assays, results were calculated using a unique cut-off level (that giving results most comparable to those obtained with the Clearview Simplify D-dimer assay, mentioned above and presented in Table 2). As shown in Table 3, the risk of VTE recurrence was significantly higher in patients with abnormal vs. normal D-dimer levels according to both the qualitative and all the quantitative D-dimer assays only among the patients
70 years old. Furthermore, the percentage of patients with abnormal D-dimer levels in both qualitative and quantitative D-dimer assays was significantly higher (2- to 3-fold) in patients aged over 70 than in those
70 years old.
|
70 and > 70 years. For each assay the results calculated using the lowest cut-off level giving a statistically significant hazard ratio are reported (patients aged
70 years: VIDAS D-dimer Exclusion = 600 ng/mL, Innovance D-DIMER = 500 ng/mL, HemosIL D-dimer HS = 250 ng/mL, and STA Liatest D-dimer=700 ng/mL; patients aged >70 years: VIDAS D-dimer Exclusion=1200 ng/mL; Innovance D-DIMER=900 ng/mL; HemosIL D-dimer HS = 450 ng/mL; STA Liatest D-dimer = 1000 ng/mL). As shown in Table 4, to obtain statistically significant hazard ratios higher cut-off levels should be used for elderly patients.
|
| Discussion |
|---|
|
|
|---|
Recently, the PROLONG study13 evaluated the role of D-dimer levels in tailoring the duration of treatment with a vitamin K antagonist in patients with a previous unprovoked VTE episode. The results of that study indicate that patients with altered D-dimer levels, assessed 1 month after withdrawal of vitamin K antagonism, have a significantly higher risk of VTE recurrence and benefit from prolonged treatment with a vitamin K antagonist. As PROLONG was a multicenter study, the same D-dimer test needed to be used in all participating centers. The qualitative Clearview Simplify D-dimer assay was chosen for this purpose; this assay can be performed on citrated whole blood and does not require any instrumentation. The use of qualitative D-dimer tests is not generally recommended; observer-independent tests should be preferred in daily clinical practice in which operating conditions are less stringent than those in clinical studies. On the other hand, the choice of any other quantitative test would not have solved the question of between-assay comparability due to poor standardization of D-dimer methods. Given the poor standardization, if a D-dimer test is to be used in clinical practice to stratify the risk of VTE recurrence, assays other than that used in the PROLONG study need to be independently assessed with a similar management study. Some indications can, however, be drawn from the present study, which evaluated different quantitative D-dimer tests using frozen plasma from patients enrolled in the PROLONG study.
The present analysis shows that quantitative D-dimer assays may be useful in evaluating a patients individual risk of VTE recurrence following a first unprovoked event. These assays seem particularly advantageous since they allow the selection of different cut-off levels according to a patients characteristics. We found that when the cut-off levels validated for use in diagnostic strategies for exclusion of VTE in symptomatic patients were used, even when assessing VTE recurrence risk, the percentage of subjects with abnormal D-dimer results was higher than that recorded in the qualitative assay; as a consequence, the rate of VTE recurrence in subjects with abnormal D-dimer levels was lower than that recorded in the PROLONG study. These data suggest that slightly higher cut-off levels than those used in diagnostic strategies used to rule out VTE (usually around 500 ng/mL for D-dimer assays whose results are expressed in fibrinogen equivalent units and around 250 ng/mL for those whose results are expressed in D-dimer units)24 may be better suited for predicting the risk of recurrence.
As shown in Figures 1 and 2, when cut-off levels were increased, the percentage of subjects with altered results decreased and the rate of recurrence increased; in contrast, the rate of recurrence in subjects with normal D-dimer levels remained substantially unchanged. It seems, therefore, that selection of a progressively higher D-dimer cut-off does not translate into a lower negative predictive value. Our interpretation of these data is that an abnormal D-dimer level is a stronger positive predictor for the risk of VTE recurrence than a normal D-dimer level is a negative predictor.
It is well known that D-dimer levels increase with age, particularly in the presence of co-existing functional impairment, because of a combination of factors, including reduced renal clearance, increased fibrinogen levels and the presence of occult disease.25,26 It has been shown that in healthy subjects D-dimer levels are approximately four times higher in the highest age quartile.25 As a result, it has been reported that the specificity and hence diagnostic utility of D-dimer measurements for excluding VTE is lower in older patients.27–29 In one study, it was shown that the specificity of the VIDAS D-dimer assay improved when a higher cut-off level was used in elderly patients without any decrease in sensitivity.30 The PROLONG study showed that abnormal Clearview Simplify D-dimer results were significantly associated with an increased risk of VTE recurrence; however, analyzing the results according to the age of the patients, it was found that the association did not reach statistical significance in patients aged >70 years. Similar results have been obtained for quantitative assays when a single cut-off level is used for all patients. A theoretical advantage of using quantitative D-dimer tests is that different thresholds discriminating between abnormal and normal can be used according to the patients characteristics. The present results support the use of different cut-off levels depending on a patients age. Indeed, using a higher cut-off value, abnormal D-dimer results in the study were associated with a significant increase in the risk of VTE recurrence even in elderly patients.
Some important limitations of the present study need to be pointed out. First and foremost, this was a post-hoc analysis of a prospective study. It is likely that the hazard ratios calculated for the quantitative assays were exaggerated since cut-off levels were determined knowing which subjects had or had not had recurrent VTE. Second, not all the patients enrolled in the PRO-LONG study could be evaluated in the present analysis because of the unavailability of frozen plasma aliquots and the exclusion of those patients who had been randomized to resume treatment with a vitamin K antagonist. The number of patients examined was, therefore, lower than that in the PROLONG study; in particular the number of recurrent VTE events was low, thus affecting the accuracy of the findings and widening the confidence intervals of estimates. This limitation is even greater when different subgroups of patients were analyzed according to age. Finally, each D-dimer assay should be evaluated in specifically designed management studies to prospectively validate their use in assessing the risk of VTE recurrence.31 We do, however, believe that the present results can be of some use for further prospective studies.
| Appendix |
|---|
|
|
|---|
U.O. Angiologia e Malattie della Coagulazione "Marino Golinelli", Policlinico S.Orsola-Malpighi, Bologna – G. Palareti (n=112, Coordinating Center); I° Div. di Medicina Interna - Centro Emostasi e Trombosi, Arcispedale Santa Maria Nuova, Reggio-Emilia – A. Ghirarduzzi (n=51); A.O. Istituti Ospitalieri, Cremona – S. Testa (n=48); Centro Regionale Malattie Emorragiche e Trombotiche - Div. Ematologia, Ospedale S. Bortolo, Vicenza – A. Tosetto (n=21); Sezione Trasfusionale, Ospedale San Leopoldo Mandic, Merate – N. Erba (n=15); Servizio di Prevenzione e Terapia della Trombosi, Ospedale "Ex Busonera", Padova – V. Pengo (n=15); Centro Sorveglianza Anticoagulati - Malattie della Coagulazione e Angiologia Medica - Div. di Med. Interna, Ospedale "S. Cuore di Gesu'", Gallipoli – L. Ria (n=13); Centro Emostasi, Ospedale Regionale, Parma – C. Pattacini (n=10); Laboratorio Analisi - Ambulatorio per il Controllo della Terapia Anticoagulante Orale, Presidio Ospedaliero di Faenza – E. Bucherini (n=9); Laboratorio Analisi - Divisione di Cardiologia, Ospedale di Bentivoglio – E. Cerè (n=9); Laboratorio di Patologia Clinica, Presidio Ospedaliero S.Maria Incoronata dell'Olmo, Cava dei Tirreni – C. Villani (n=8); Centro Trombosi, A.O. di Careggi Universita' di Firenze – D. Prisco (n=4); U.O. Medicina Interna, Policlinico Universitario Messina – A. Trifiletti (n=4); Centro per lo Studio delle Coagulopatie a Rischio Trombotico, Osp. Galliera, Genova – A. Schenone (n=2).
| Acknowledgemnts |
|---|
|
|
|---|
| Footnotes |
|---|
Authorship and Disclosures CL analyzed the data and wrote the paper; GP designed and promoted the study; BC was responsible for the enrollment of the patients and follow-up; MC performed the D-dimer assays; ATo and ATr contributed to data interpretation and helped revise the manuscript; all authors checked the final version of the manuscript. CL, GP, BC, ATo and ATr received lecture fees from Instrumentation Laboratory. CL and GP received lecture fees from Roche Diagnostics. There are no other actual or potential conflicts of interest relevant to this article.
Received for publication September 28, 2007. Revision received January 7, 2008. Accepted for publication February 1, 2008.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. S. Adam, N. S. Key, and C. S. Greenberg D-dimer antigen: current concepts and future prospects Blood, March 26, 2009; 113(13): 2878 - 2887. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||