Disorders of Iron Metabolism |
1 Department of Clinical Chemistry, Radboud University Nijmegen Medical Centre, Nijmegen, The Netherlands
2 Pharmaceutical Sciences Division, Kings College London, London, UK
3 Department of Clinical Chemistry, Imperial College HealthCare NHS Trust, Hammersmith Hospital Campus, London, UK
4 Department of Clinical and Experimental Medicine, Section of Internal Medicine, University of Verona, Verona, Italy
5 Laboratory of Molecular Biology & Immunobiotechnology, Department of Biochemistry, Hellenic Pasteur Institute, Athens, Greece
6 Intrinsic Life Sciences, La Jolla, CA, USA
7 Division of Advanced Medicine, Medical Research Institute/Division of Nephrology, Kanazawa Medical University, Ishikawa, Japan
8 CRUK Institute for Cancer Studies, University of Birmingham, Birmingham, UK
9 Department of Medicine, David Geffen School of Medicine at UCLA, LA, USA
10 Department of Epidemiology and Biostatistics, Radboud University Nijmegen Medical Centre, Nijmegen, The Netherlands
Correspondence: Dorine W. Swinkels, Department of Clinical Chemistry, Radboud University Nijmegen Medical Centre, P.O. Box 9101 6500 HB, Nijmegen, The Netherlands. E-Mail: D.Swinkels{at}akc.umcn.nl
|
|
|---|
Key words: hepcidin, iron, quality control.
|
|
|---|
|
|
|---|
Specimens
Eight urine samples with a wide range of hepcidin concentrations were collected from healthy subjects (samples 1–5) and patients (samples 6–8) (October 2007) with informed consent according to the declaration of Helsinki. Synthetic hepcidin-25 (Peptide Institute Inc., Osaka, Japan, net hepcidin peptide weight is precisely determined by amino acid analysis after acid hydrolysis) was added to a final concentration of 13.1 nmol/mmol creatinine to a urine sample that by SELDI-TOF MS7 was found to have a hepcidin concentration below the lower limit of detection of 0.5 nmol/L (sample 5). Seven plasma pools were composed from hospitalized patient sample remnants (March 2008), so as to cover a wide variation in hepcidin levels (samples 9–15). All samples were centrifuged for 10 min at 2600 g, and immediately stored in aliquots at –80°C. Two weeks after collection and storage, the samples were shipped on dry ice to all participants, and measured within four weeks of receipt, except for method VI (see below), that was carried out in July 2008. All samples underwent only one freeze-thaw cycle before analysis to minimize changes that may differently affect the various methods, among which are the formation of hepcidin aggregates and breakdown products.
Hepcidin methods
Characteristics of the methods used for the plasma and urine hepcidin measurements of the present study are schematically presented in Table 1.
|
View this table: [in a new window] [Download PPT slide] |
Table 1. Characteristics of methods used for plasma and urine hepcidin measurements.
|
Method II is also based on SELDI-TOF MS. Hepcidin was first extracted from the sample using Macro-Prep® CM Support beads (Bio-Rad Laboratories). The extract was applied to NP20 chips (Bio-Rad Laboratories). Synthetic hepcidin-24 peptide (custom made, Peptide Institute Inc.) was used as an internal standard.
For method III, urine samples were de-salted using C8 Clinprot beads (Bruker Daltonik) and analyzed by MALDI-TOF MS. Plasma samples were assayed by SELDI-TOF-MS using Cu2+ loaded IMAC chips. Synthetic human hepcidin-25 (Peptide Institute Inc.) was used as an external standard in both assays.
Method IV is based on LC-MS/MS. [15N,13C2]Gly12,20 hepcidin (heavy hepcidin) was added to plasma or urine as internal standard. Magnetic nanoparticles (Bruker Daltonik) were used to extract hepcidin from the samples followed by LC-MS/MS analysis using selected reaction monitoring of the triple charged precursor fragmenting to the double charged product ion for both heavy hepcidin and hepcidin.
Method V is a competitive ELISA. 96-well plates were coated with in house-prepared anti-human hepcidin antibody and biotinylated hepcidin-25 as tracer. Custom synthesized hepcidin-25 was used as an external standard.
Method VI is a competitive RIA using a 125I labeled synthetic hepcidin-25 (Bachem) with an in-house rabbit anti-hepcidin polyclonal antibody (against hepcidin-KLH conjugate) using a secondary antibody-PEG assisted separation. Synthetic hepcidin-25 (Bachem) was used as an external standard.
Method VII is based on LC-MS/MS. Isotopic human synthetic hepcidin-25 (Peptide Institute Inc.) was used as an internal standard.
Method VIII is a competitive ELISA: plates were coated with an in -house prepared hepcidin-25-His peptide and a polyclonal antibody against recombinant (in-house prepared) hepcidin-25-His was used to establish competition. Horseradish peroxidase labeled anti-rabbit antibody was used as secondary antibody. Hepcidin-25-His was used as an external standard.
Statistical methods
The study was designed to compare hepcidin levels as well as the repeatability of the methods used for serum and urine, respectively. With respect to the repeatability, the magnitude of variation that exists between samples and between measurements of the same sample relative to the total variation are of interest. Accordingly, we partitioned the total variance of each hepcidin method into the following components: i) the between-sample variance and ii) the analytical variance. Our design allowed us to divide the latter into three subcomponents: the between-day variance, the between triplicate variance and the residual analytical variance, i.e. the part of the analytical variance that cannot be attributed to the other two.
A linear mixed model was used to estimate these variance components of each method separately. The dependent variable was hepcidin outcome, and the independent random variables were: sample (plasma: 7 levels, urine: 8 levels), day (4 levels) and repeated measurement (3 levels). We found that the estimated percentage variance due to the triplicate measurements was very small and considerably smaller than the variance due to the between-day variation. Consequently, this term was omitted from the final model and not presented separately. The SD (absolute error), the CV (relative error) and the percentage variance relative to the total variance of each random variable are presented and for each method separately.
|
|
|---|
|
View this table: [in a new window] [Download PPT slide] |
Table 2. Mean (±SD) urine and plasma hepcidin levels presented by sample ID and method.
|
In general, differences in hepcidin levels between methods might be due to: (i) the use of different calibration solutions with level assignments based on different techniques; (ii) to hepcidin aggregation of either the standard solution or the sample; or (iii) hepcidin binding to
2 macroglobuline or albumine16 or (iv) the existence of three hepcidin isoforms hepcidin-25, 22 and 20. These four points may differentially affect IC and MS measurements, and urine and serum quantifications. More specifically, it was recently found that around 90% of the circulating hepcidin is bound to
2 macroglobulin in the blood. These observations not only raise the question whether we should measure total, bound or unbound hepcidin, but also what the methods evaluated actually measure. The search for an answer to these questions is a new challenge for which much can be learned from the measurements of steroid and thyroid hormones. Another cause for differences between IC and MS methods is that IC methods lack the selectivity to distinguish hepcidin-25 from hepcidins-20 and -22. However, the implications of including the latter two isoforms in the assay on the total hepcidin values reported in the various iron disorders are not yet known.
Most methods are similar in both analytical variation and between-sample variation (Table 3). Of note is that the between-sample CV is lower for plasma than for urine hepcidin. This might indicate that the difference between urine and plasma is not due to the method, but more to biological mechanism, e.g the hepcidin excretion pathways. MS methodologies II, IV and VII, exploiting an internal standard, show slightly lower contribution of the analytical variance to the total variance compared to the other MS methods I and III. These findings corroborate the assumption that the use of an internal standard decreases the analytical variation of the MS-techniques. IC-method V shows a high between-sample CV in combination with a relatively high analytical CV. Furthermore, IC-methods VI and VIII express the lowest between-sample CV compared to all other methods. However, both methods also display low analytical variation. This relatively low between-sample variation of both the latter IC-methods illustrates the difficulties in the generation of specific antibodies for hepcidin.
|
View this table: [in a new window] [Download PPT slide] |
Table 3. Sample means and variations by method for urine and plasma hepcidin levels.
|
Spearman correlations between the individual sample mean hepcidin values obtained by most methods were generally high (range 0.62–1.00), except for correlations with method VIII that are somewhat disappointing (range 0.04–0.18). This should be interpreted with caution due to the small number of samples analyzed (n=7 and n=8 for plasma and urine, respectively).
In summary, hepcidin levels reported by the various methods vary considerably but analytical variance is generally low and similar for all methods. We recommend further harmonization of the various hepcidin assays by: (i) introducing an internal standard for all the MS-based methods used for clinical studies; (ii) reaching consensus on level assignment and level adjustment of the calibrators used in every procedure; (iii) production of a calibrator that mimics patient sera and (iv) regular testing of shared samples and/or calibrators that are commutable and have been value assigned for quality control.
JJCK and EHJMK designed research, performed statistical analysis, contributed analytical tools, interpreted data and wrote the paper. SSB, MB, NC, DG, RCH, VK, AM, GO, NT, CT, DGW and TG contributed analytical tools and edited the paper. JCMH, designed and performed statistical analysis, interpreted data and co-wrote the paper.
DWS designed research and statistical analysis, interpreted the data, wrote the paper and co-ordinated the study. DWS and HT steer the www.hepcidinanalysis.com initiative that serves the scientific community with hepcidin analysis in body fluids.
VK and AM have patents applications to disclose (ELISA hepcidin); GO and TG are employees of Intrinsic Life Sciences and have ownership of stocks to disclose. The other authors have no conflict of interests to declare.
Received for publication April 18, 2009. Revision received May 28, 2009. Accepted for publication June 22, 2009.
|
|
|---|
}-2-macroglobulin in blood. Blood 2009;113:6225–36.This article has been cited by other articles:
![]() |
G. Bergamaschi and L. Villani Serum hepcidin: a novel diagnostic tool in disorders of iron metabolism Haematologica, December 1, 2009; 94(12): 1631 - 1633. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||